YSI 600DW-B Sonde User Manual

Page 68

Advertising
background image

Sondes

Section

2

Select

1 – Start sampling option to start discrete sampling. After the initial sampling time interval has

passed (4 seconds in the screen above), sequential lines of data will appear on the screen.

=============================================================================

Date Time Temp SpCond Sal Cl2 Depth pH Turbid Battery

mm/dd/yy hh:mm:ss C mS/cm ppt mg/L feet NTU volts
------------------------------------------------------------------------------
*** 1-LOG last sample 2-LOG ON/OFF, 3-Clean optics***
05/05/97 10:04:40 21.57 0.009 0.00 0.95 -0.293 7.55

0.3

10.2

05/05/97 10:04:44 21.57 0.009 0.00 0.96 -0.300 7.55

0.3

10.3

*** LOG is ON, hit 2 to turn it OFF, 3-Clean optics***
05/05/97 10:04:48 21.58 0.009 0.00 0.95 -0.302 7.55

0.3

10.3

05/05/97 10:04:52 21.58 0.009 0.00 0.95 -0.302 7.55

0.3

10.3

05/05/97 10:04:56 21.58 0.009 0.00 0.95 -0.303 7.55

0.3

10.2

05/05/97 10:05:00 21.58 0.009 0.00 0.96 -0.303 7.55

0.3

10.3

*** 1-LOG last sample 2-LOG ON/OFF , 3-Clean optics***
05/05/97 10:05:04 21.58 0.009 0.00 0.96 -0.304 7.55

0.3

10.3

Sample logged.
05/05/97 10:05:08 21.58 0.009 0.00 0.95 -0.305 7.55

0.3

10.3

The following prompt will appear just below the screen header:
1-LOG last sample, 2-LOG ON/OFF, 3-Clean optics

• By entering 1–LOG last sample, A single line of data can be logged to sonde memory and the

following message will be displayed:

Sample logged.

• By entering 2-LOG ON/OFF, a set of data can be logged to memory and the following message will

be displayed:

LOG is ON, hit 2 to turn it OFF, 3-Clean optics. Press 2 again to terminate logging.

• By entering 3-Clean optics, if your unit has an optical (turbidity, chlorophyll or rhodamine WT)

probe, the wiper will clean the optical surface

. The 3-Clean optics portion of the prompt will only

appear if an optical probe was installed and enabled.


Select

Esc or press 0 and return to the Discrete Sampling menu.

From the Discrete Sampling menu, select

2-Sample Interval to type a number that represents the number

of seconds between samples. The maximum sample interval is 32767 seconds (9+ hours). The factory
default sample interval is 4 seconds and works best for most discrete sampling applications. The shortest
possible sample interval for obtaining new free chlorine readings is 4 seconds. If a smaller interval is
selected, then the Cl2 readings remain constant for the number of samples necessary to fill 4 seconds.

NOTE: If your have used your sonde with a 650 MDS data logger, the Sample Interval automatically be
changed to 0.5 seconds.










YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-61