Calibration example – YSI 600DW-B Sonde User Manual

Page 37

Advertising
background image

Sondes

Section

2

be used during calibration if you press

Enter without inputting another value. Be sure not to

accept default values unless you have assured that they are correct. If no default value appears,
you must type a numerical value and press

Enter.


4.

After you input the calibration value, or accept the default, press

Enter. A real-time display will

appear on the screen. Carefully observe the stabilization of the readings of the parameter that is
being calibrated. When the readings have been stable for approximately 30 seconds, press

Enter

to accept the calibration. The calibrated value is bolded on the example screen on the following
page.


5. Press

Enter to return to the Calibrate menu, and proceed to the next calibration.


CALIBRATION EXAMPLE


The example below for calibration of specific conductance, is designed to demonstrate the general calibration
protocol for all parameters. From the Calibrate menu, press

1-Conductivity and the following display will be

shown.






Select

1-SpCond (the generally-recommended method for calibration of a conductivity sensor), the following

prompt will appear which calls for your numerical input.



As noted above, the number in parentheses is the default value of this parameter and will be used in the
calibration if only

Enter is pressed without typing in another value. Similar prompts will be displayed during

the calibration of all parameters, but for some sensors, such as pH, no default values are provided. In these
cases, the user must input a numerical value and then press

Enter.

After the calibration value is input and

Enter is pressed, a real-time display similar to the following will then

appear on the screen..

------------Cond calibration-----------
1-SpCond
2-Cond
3-Salinity

Select option (0 for previous menu): 1

Enter SpCond in mS/cm (10):

Date Time Temp SpCond Sal Cl2 Cl2 chrg pH Battery
mm/dd/yy hh:mm:ss C mS/cm ppt mg/L volts
------------------------------------------------------------------------------
To calibrate, press <Enter> when the readings are stable.

05/05/97 08:39:51 20.83 9.602 8.837 1.04 8.5 7.06 10.2


Note that all parameters that have been enabled will appear - not just the one being calibrated at the moment.
The user should carefully observe the stabilization of the readings of the parameter that is being calibrated

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-30