YSI 600DW-B Sonde User Manual

Page 53

Advertising
background image

Sondes

Section

2






















Select

1-Directory to view all files currently stored in the sondes flash disk memory, the screen below

shows 6 files. All data files (.dat extension) could be uploaded to EcoWatch for viewing or plotting, but
you do not need to upload all files in the directory. The file with the .glp extension contains the calibration
record of the sonde. It is fully described in Sections 2.9.2 and 2.9.3 below. Details of any of the studies
can be obtained by pressing the number key associated with the file in the Directory.



Select

2-Upload to view file lists in memory, and upload the data to PC-based software.


Prior to upload, a “

Time window” display appears that will allow you to select portions of the logged data

to upload. You may select

1-Proceed to upload all logged data from the dates and times displayed.

Filename Samples

1-BRIDGE1.dat 19
2-BRIDGE2.dat 27
3-UPLAKE.dat 33
4-CLRLAKE2.dat 167
5-DWNLAKE2.dat 31
6-00003001.glp 3

Select option (0 for previous menu): 4


-----------------File-----------------
-
1-Directory 4-View file
2-Upload 5-Quick view file
3-Quick Upload 6-Delete all files

Select option (0 for previous menu): 1

-------------File details--------
------
1-View file
2-File:CLRLAKE2.dat
3-Samples: 167
4-Bytes: 4421
5-First:08/23/2001
6-First:08:33:40
7-Last :08/23/2001
8-Last :09:04:20
9-Interval:00:00:10
A-Site:Clear Lake

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-46

Advertising
This manual is related to the following products: