2 sonde communication problems – YSI 600DW-B Sonde User Manual

Page 241

Advertising
background image

Troubleshooting

Section

6

The following troubleshooting tables can help you if you encounter problems with software,
communication protocol, or sensor malfunctions other than calibration errors.

6.2 SONDE COMMUNICATION PROBLEMS


SYMPTOM POSSIBLE

CAUSE ACTION

REF

Sonde not powered

Check 12 vdc source

7.2

Cable connection is loose

Check both ends of cable; secure
connectors

Damaged connectors

Check pins at both ends, insure
they are straight, dry and clean.

Cannot communicate
with sonde

Com port not selected

Change to other com port, other
peripheral on the same port
(Internal mouse). Try other PC,
650 display/logger or dumb
terminal

2.4.3

Unmatched baud rate between
host and sonde

Match the baud rate

2.4.3

Host is too slow

Use faster computer

Interface cable failure

Check cable for damage. If
necessary, return for service

7

Internal failure

Return sonde for service

7