Conductivity – YSI 600DW-B Sonde User Manual

Page 38

Advertising
background image

Sondes

Section

2

and, when the readings are stable for approximately 30 seconds, press

Enter to implement the calibration and

the following message will appear.

Calibrated. Press <Enter> to continue.

NOTE: If an ERROR message appears, begin the calibration procedure again. Be certain that the value you
enter for the calibration standard is correct. Also see Section 8, Troubleshooting for more information on
error messages.


CAUTION:
Be certain to immerse the entire sonde in solution standards for calibration of all
parameters. Most calibrations require readings not only from the sensor being calibrated but also
from the temperature sensor.

Specific start-up calibration procedures for all sensors that commonly require calibration are
provided in the following paragraphs of this section. Remember that these are basic protocols
designed to get the user up and running with your 6-series sonde. A more-detailed discussion of
sensor calibration can be found in Section 2.9.2.


CONDUCTIVITY

This procedure calibrates conductivity, specific conductance, salinity, and total dissolved solids.

Place the correct amount (see Tables 1-4) of 10 mS/cm conductivity standard (YSI 3163 is recommended)
into a clean, dry or pre-rinsed calibration cup.

Before proceeding insure that the sensor is as dry as possible. Ideally, rinse the conductivity sensor with a
small amount of standard that can be discarded. Be certain that you avoid cross-contamination of standard
solutions with other solutions. Make certain that there are no salt deposits around the pH/ORP probe,
particularly if you are employing standards of low conductivity.

Carefully immerse the probe end of the sonde into the solution. Gently rotate and/or move the sonde up
and down to remove any bubbles from the conductivity cell. The probe must be completely immersed past
its vent hole. Using the recommended volumes from the table in the previous subsection should insure that
the vent hole is covered.

Allow at least one minute for temperature equilibration before proceeding.

From the Calibrate menu, select

Conductivity to access the Conductivity calibration procedure and then 1-

SpCond to access the specific conductance calibration procedure. Enter the calibration value of the
standard you are using (mS/cm at 25

°C) and press Enter. The current values of all enabled sensors will

appear on the screen and will change with time as they stabilize.

Observe the readings under Specific Conductance or Conductivity and when they show no significant
change for approximately 30 seconds, press

Enter. The screen will indicate that the calibration has been

accepted and prompt you to press

Enter again to return to the Calibrate menu.


Rinse the sonde in tap or purified water and dry the sonde.

NOTE: The YSI conductivity system is very linear over its entire 0-100 mS/cm range. Therefore, it is
usually not necessary to use calibration solutions other than the 10 mS/cm reagent recommended above for
all environmental applications from low conductivity freshwater to seawater. YSI does offer the 3161

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-31

Advertising
This manual is related to the following products: