6 zero adjustment, Zero adjustment – Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 46

Advertising
background image

Operation

pHotoFlex / pHotoFlex Turb

46

ba75491d04

07/2006

4.5.6

Zero adjustment

The zero adjustment, i. e. measuring and storing the absorbance of a
cell filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the ambi-
ent temperature has changed.

Only perform the zero adjustment against distilled water in an optically
perfect cell. The zero adjustment must be performed separately for
each cell type.

5

Clean the cell (see section 5.2.2).

6

Insert the cell (see section 4.2).

7

Start the measurement with <START/ENTER>.

The measurement result is displayed when the measurement
is finished.

1

Press the <M> key (long pressure) repeatedly until the Pho-
tometry
measuring mode is selected.

2

Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Concentration is selected.

3

Press the <CAL/ZERO> key.

The menu for adjustment measurements opens up.

Photometry \ Absorbance
i Insert sample

i Start measurement
with <START>

16 mm

610 nm

01.02.05 15:12

Photometry \ Absorbance

0.532

16 mm

610 nm

01.02.05 15:12