2 resetting the photometer settings, 3 resetting the ph settings, Ee section 4.10.3) – Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 77

Advertising
background image

pHotoFlex / pHotoFlex Turb

Operation

77

ba75491d04

07/2006

4.10.2 Resetting the photometer settings

With the Photometry / Reset function, all photometer settings are reset.

4.10.3 Resetting the pH settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

The following settings for pH measurements are reset to the default
settings with the Reset function:

Setting

Default settings

PROG

0

Cell size

16 mm

Measured parameter

Konzentration

Wavelength

436 nm

Blank values

all erased

Setting

Default settings

Cal. type

TEC

Calibration interval

7 d

Measured parameter

pH & ORP

Asymmetry (Asy)

0 mV

Slope (Slp.)

-59.16 mV/pH

Temperature, manual

25 °C

Temperature

°C