Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 47

Advertising
background image

pHotoFlex / pHotoFlex Turb

Operation

47

ba75491d04

07/2006

Note
If

i Calibration error! was displayed as the calibration result, a note au-

tomatically reminds you of another zero adjustment before the next
measurement.
Measuring is not possible without a valid zero adjustment.

4

Using <

> <> and <START/ENTER>, select and start the

Zero adjust. function.

The menu-guided zero adjustment starts.
Follow the instructions on the display.

5

Insert the cell (see section 4.2)

6

Set another cell with <

> <> and <START/ENTER> as nec-

essary.

7

Start the measurement of the zero adjustment with <START/
ENTER>
.

After measuring, the result of the zero adjustment is displayed
and stored.

i Zero adjust. successful! (successful zero adjustment) or

i Calibration error! (zero adjustment not successful)
is displayed as the result.
The zero adjustment is completed.

8

Confirm the result with <START/ENTER>.

The zero adjustment is completed.
The meter is ready to measure.

Zero adjust.

Measure blank value

Delete blank value

Photometry \ Adjustment

i Insert zero
cell (dist. water)

Cell

16 mm

Start measurement

Zero adjust.