pHotoFlex / pHotoFlex Turb

Operation

79

ba75491d04

07/2006

4.13

Administrating user-defined methods

User-defined programs can be

! entered
! read out
! erased.

To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex (Turb).

Administration of the method data can be carried out in the LSdata soft-
ware (see operating manual LSdata) or with the aid of a terminal pro-
gram (section 4.13.1).

You need the following data in any case:

Data

Possible entries/examples

Program No:

900 ... 999

Model:

Any name
(max. 5 characters)

Cuvette [mm]:

16 or 28

Wave length [nm]:

436, 517, 557, 594, 610 or 690

MRB:
(measuring range beginning)

e.g. 0.05

MRE:
(measuring range end)

e.g. 8.00

E0: (Offset)

e.g. 0.0

Slope: e.g.

1.0

Resolution:

0.0001, 0.001, 0.01, 0.1, 1, 2, 5, 10
or 100

Formula: (citation form)

e.g. PO4-P (max. 9 characters)

Unit

e.g. mg/l (max. 7 characters)