Administrating user-defined methods
User-defined programs can be
! entered
! read out
! erased.
To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex (Turb).
Administration of the method data can be carried out in the LSdata soft-
ware (see operating manual LSdata) or with the aid of a terminal pro-
gram (section 4.13.1).
You need the following data in any case:
Data
Possible entries/examples
Program No:
900 ... 999
Model:
Any name
(max. 5 characters)
Cuvette [mm]:
16 or 28
Wave length [nm]:
436, 517, 557, 594, 610 or 690
MRB:
(measuring range beginning)
e.g. 0.05
MRE:
(measuring range end)
e.g. 8.00
E0: (Offset)
e.g. 0.0
Slope: e.g.
1.0
Resolution:
0.0001, 0.001, 0.01, 0.1, 1, 2, 5, 10
or 100
Formula: (citation form)
e.g. PO4-P (max. 9 characters)
Unit
e.g. mg/l (max. 7 characters)