Operation

pHotoFlex / pHotoFlex Turb

66

ba75491d04

07/2006

Note
Before measuring the third calibration standard of 0.02 NTU/FNU you
can exit the calibration with <ESC> at any time.
The new calibration data is discarded. The old calibration data is further
used.

4

Align the cell:
! Marked cell:

– Align the marking on the cell cap with the marking on the

cell shaft.

– Press and time hold the <START/ENTER> key until the

measured value is displayed.

! Unmarked cell (see page 62)

– Press and hold the <START/ENTER> key.

– Slowly and in small steps turn the cell by one complete

rotation (by 360 °).

– After each step wait for a short time until the displayed

measured value is stable.

– Turn the cell back to the position with the lowest mea-

sured value.

5

Release the <START/ENTER> key.

Measurement of the calibration standard begins.

6

Repeat the steps 4 - 6 with the calibration standards, 10.0 NTU/
FNU and 0.02 NTU/FNU.

After measuring the 0.02 NTU/FNU calibration standard, the
calibration result (

i Calibration successful! or i Calibration er-

ror!) is displayed and stored.
Calibration is completed.

7

Confirm the calibration result with <START/ENTER>.

The display shows instructions for the first measurement.

i

Turb.

= 1000 FNU/NTU

i Start calibration
by releasing <START>

Turbidity \ Calibration