3 ph value / orp voltage, 4 turbidity, Ph value / orp voltage – Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 94: Turbidity

Advertising
background image

Technical data

pHotoFlex / pHotoFlex Turb

94

ba75491d04

07/2006

7.3

pH value / ORP voltage

7.4

Turbidity

Measuring ranges,

resolution

Variable

Measuring range

Resolution

pH

- 2.00 ... + 16.00

0.01

U [mV]

- 1000 ... + 1000

1

T [°C]

- 5.0 ... + 100.0

0.1

T [°F]

- 23.0 ... + 212.0

0.1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 20 ... + 100

1

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

pH *

± 0.01

+ 15 °C ... + 35 °C

U [mV]

± 1

+ 15 °C ... + 35 °C

T [°C]

± 0.3

0 °C ... + 55 °C

T [°F]

± 0.54

0 °C ... + 55 °C

* when measuring in a range of ± 2 pH around a calibration point

Measuring principle

Nephelometric measurement according to DIN EN ISO 7027

Light source

Infrared LED

Measuring range

0.01 ... 1100 NTU/FNU

Resolution

in the range
0.01 ... 9.99

max 0.01 NTU/FNU

in the range
10.0 ... 99.9

max 0.1 NTU/FNU

in the range
100 ... 1100

max 1 NTU/FNU

Accuracy

in the range
0 ... 1000 NTU/FNU

± 2% of the measured value

or ± 0.01 NTU/FNU

Measuring time

4 seconds

Calibration

Automatic 3-point calibration