Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 61

Advertising
background image

pHotoFlex / pHotoFlex Turb

Operation

61

ba75491d04

07/2006

Continuing for three-

point calibration

(Cal. type TEC)

14

Thoroughly rinse the electrode with distilled water.

15

Immerse the electrode in buffer solution 3.

16

If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <

> <> and <START/ENTER>

in the Set temperature setting.

17

Using <

> <>, select Continue and press <START/EN-

TER>. The buffer is measured.

The measured value is checked for stability (AutoRead).

18

Wait for the end of the AutoRead measurement or Terminate
AutoRead
with <START/ENTER> and take over the calibration
value.

The new calibration values are displayed as an informative
message (

i

).

You have the following options:
! Accept the new calibration values with <START/ENTER>.

Subsequently, the calibration record is displayed and output
to the interface at the same time.

! To switch to the measured value display without accepting

the new calibration values, press <M> (short pressure) or
<ESC>.

i Buffer value = 4.01
i U = 184 mV
i Temperature =

24.8 °C

Terminate AutoRead

pH & ORP \ Calibration