pHotoFlex / pHotoFlex Turb

64

ba75491d04

07/2006

Note
To keep the drift as low as possible, the time for aligning the cell while
pressing and holding the <START/ENTER> key is limited to 30 sec-
onds. After this time, the meter automatically starts measuring or cali-
brating.

Display if the

measuring range is

exceeded

If the measured value is outside the measuring range of the
pHotoFlex Turb, it is indicated on the display:

7

Align the cell:
! Marked cell

– Align the marking on the cell cap with the marking on the

cell shaft.

– Press and for a short time hold the <START/ENTER>

key until the measured value is displayed.

! Unmarked cell (see page 62)

– Press and hold the <START/ENTER> key.

– Slowly and in small steps turn the cell by one complete

rotation (by 360 °). After each step wait for a short time
until the displayed measured value is stable.

– Turn the cell back to the position with the lowest mea-

sured value.

8

Release the <START/ENTER> key.

Measurement starts. The measured value is displayed.

9

Repeat the steps 2 to 8 for further samples.

Turbidity

157.0

FNU
NTU

01.02.05 15:12

Turbidity

> 1100

FNU
NTU

01.02.05 15:12