What to do if...

pHotoFlex / pHotoFlex Turb

90

ba75491d04

07/2006

Obviously incorrect

measured values

Sensor symbol flashes

6.4

Turbidity

Error message

Measured values

obviously incorrect

Measured value display

< 0.01 FNU

Cause

Remedy

pH electrode:

– Not connected

– Connect electrode

– Cable broken

– Replace cable or electrode

– pH electrode unsuitable

– Use suitable electrode

– Temperature difference

between buffer and test
sample too high

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Cell not correctly inserted

– Lock cell into place

– Cell contaminated

– Clean the cell

– Calibration too old

– Carry out calibration

Cause

Remedy

– Measured value outside the

measuring range

– not possible