pHotoFlex / pHotoFlex Turb

Operation

57

ba75491d04

07/2006

Sample printout of a

record

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration. The
asymmetry and slope are evaluated separately. The worse evaluation
of both is taken into account. The evaluation appears on the display
and in the calibration record.

31.10.03 16:13
pHotoFlex Ser. no. 12345678
Calibration pH & ORP
Calibration date

31.10.03 16:13:33

Calibration interval

7 d

AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope

-60.2 mV/pH

Asymmetry 4.0 mV
Sensor +++

Sample dis-
play

Calibration
record

Asymmetry
[mV]

Slope
[mV/pH]

+++

-15 ... +15

-60.5 ... -58

++

-20 ... +20

-58 ... -57

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the electrode according
to the electrode operating
manual

----

----

< -30 or
> 30

... -62 or
... -50

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...