2 photometry, 3 ph value / orp voltage, Photometry – Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 88: Ph value / orp voltage

Advertising
background image

What to do if...

pHotoFlex / pHotoFlex Turb

88

ba75491d04

07/2006

6.2

Photometry

Measuring range

undercut or exceeded

Obviously incorrect

measured values

6.3

pH value / ORP voltage

Measuring range

undercut or exceeded

Cause

Remedy

– Program not suitable

– Select program with suitable

measuring range

– Dilute the sample

Cause

Remedy

– Measurement disturbed by

external light

– Close the external light cover.

– Cell not correctly inserted

– Insert the cell so that it is

positioned on the bottom of
the cell shaft.

– Cell contaminated

– Clean the cell

– Cell shaft contaminated

– Clean the cell shaft

– Dilution set incorrectly

– Set the dilution

– Selected program unsuitable

– Select other program

– Zero measurement incorrect

– Perform zero measurement

– Blank value incorrect

– Remeasure the blank value

Cause

Remedy

Electrode:

– Air bubble in front of the

diaphragm

– Remove air bubble

– Air in the diaphragm

– Extract air or

moisten diaphragm

– Gel electrolyte dried out

– Replace electrode

Test sample

– The pH value lies outside the

measuring range

– not possible