pHotoFlex / pHotoFlex Turb

Operation

55

ba75491d04

07/2006

4.6.5

Calibration

Why calibrate?

pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the meter. Thus, you
should calibrate at regular intervals.

When to calibrate?

! After connecting another electrode
! When the sensor symbol flashes:

– after the calibration interval has expired

– after voltage interruption (e.g. empty batteries, empty

accumulator pack)

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
calibration.

Note
The buffers are selected in the Configuration / pH & ORP / Cal. type
menu, see section 4.6.4).

Buffer set

Name on the
display

pH values at
25 °C

Technical buffer solutions

TEC

2.00

4.01

7.00

10.01

NIST/DIN buffer solutions

NIST/DIN

1.679

4.006

6.865

9.180

12.454