Measuring the pH value
AutoRead
(Drift control)
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured pa-
rameter flashes until a stable measured value is available.
Criteria
With identical measurement conditions, the following applies:
1
Perform the preparatory activities according to section 4.6.1.
2
Immerse the pH electrode in the test sample.
3
Press the <M> key (short pressure) repeatedly until the mea-
sured variable, pH is selected.
pH & ORP \ pH
6.94
25.0
°C
[AR]
01.02.05 15:12
Measured
parameter
Reproducibility
Response time
pH value
Better than 0.01
> 30 seconds