Operation

pHotoFlex / pHotoFlex Turb

74

ba75491d04

07/2006

Output format, CSV

Data

Description

1

Date

Date of storing

2

Time

Time of storing

3

ID

adjusted ID

4

Program number

only for the measured parameter,
Concentration

5

Citation form

only for the measured parameter,
Concentration

6

AR/
BV

! AutoRead (measuring mode, pH & ORP)
! Blank value (measured parameter,

Concentration)

7

Dilution

only for the measured parameter,
Concentration

8

Measured value

! Measured value or
! Upper/lower measuring range limit

(only with measured value status, OFL/
UFL)

9

Unit of 8

! Unit of the measured value or
! Designation of dimensionless measured

values, e.g. <pH>

10

Measured value
status of 8

! VALID: Measured value valid
! INVALID: Measured value invalid
! UFL: Measured value below the lower

measuring range limit

! OFL: Measured value above the upper

measuring range limit

11

Temperature val-
ue / wavelength
of the measure-
ment

! Temperature (measuring mode, pH &

ORP)

! Wavelength (measured parameter,

Absorbance / % Transmission)

15.01.05;15:06:49;1;8;O2;[BW];[9 + 1];6.5;mmol/l;VALID;