Operating Mode: Routine BOD

Operating Mode: Routine BOD

12

BA31114e5 07/2006

BA31114e5 07/2006

13

Operating mode: Routine BOD

Sample preparation

See WTW application reports
(contained within the scope of delivery of the accessories supplied).


Screw the OxiTop

-C measuring heads onto the sample bottles

and tightly close them.

Important note:
Never use joint grease or other lubricant for your OxiTop

-C measuring

heads. Some of these products contain solvents that can cause severe damage
to the plastic housing.
The sealing of the BOD bottles is also perfectly adequate without grease.
However, you should always wipe off heavy contamination and particles on the
sealing surfaces of the rubber sleeves and OxiTop

-C. WTW accepts no

liability for damage due to the use of joint grease.


Start the measurement


Entry into the mode
“Communication with the
measuring heads“.

Preselected:
Start

(measuring head).

Use

to select the measurement range.
The filling volume required is given
in the right-hand column.


The controller stores the setting
(memory function: the last selected
measurement range is set).

Confirm the selected
measurement range for the
sample. The automatically
allocated sample number is given
in the header line (YY/MM/DD and
sequential number).

Additional information:
Type of measurement, run time, measurement range, final date, Id
number.
“Temp“ display => the AutoTemp function (see the chapter “GLP/Tools”)
is set as a default in the routine mode.

Change the Id number of the additional identification of the sample
(e.g. extraction location) as follows:

- Use

to move the cursor

to the Id number,

-

Press

and

- Use

to set the Id number required

(setting range 001 ... 255).

- Use

to confirm this.


Confirm the start of the
measurement. Contact selection:

The controller repeatedly sends the
start information in the scanner
mode until successful feedback is
received from the OxiTop®-C
measuring head.