Oxitop, C measuring head, Disposing of the batteries – Xylem OxiTopIS OC100 User Manual

Page 83

Advertising
background image

Power Supply

Power Supply

82

BA31114e5 07/2006

BA31114e5 07/2006

83

OxiTop

®

-C measuring head

Supply batteries

Batteries: 2 x lithium batteries, CR2430, e.g. WTW type Batt/OxiTop

,

order no. 209 012
These batteries ensure the energy supply of the OxiTop

®

-C measuring head.

Run time: typically 2 years
The OxiTop

®

-C measuring head does not require any batteries for data backup.


Supply batteries status messages
- are given on starting the measuring head via the OxiTop

®

OC100 controller by

display messages and a signal tone
- can be called up in the menu “GLP/Tools - Check -

info”.


Display message

Note / Meaning


Batteries LoBat !

start

The instrument can still be safely operated within the
specifications.
The running reserve of the instrument is still sufficient for
the measurement to be started.
Please obtain new batteries and replace the old ones!


Battery/ies empty!

- New

- Stop

start

The instrument can no longer be used.
The supply batteries must be replaced by new ones.
It is possible to start a new measuring head or to
complete a sample start.

Changing the supply battery

• Change the battery after the measurement has been terminated.

- Press in the snap-action holder.

- Open the upper case.

- Remove the batteries.

- Insert new batteries (2 x lithium batteries, CR2430, e.g. WTW type

Batt/OxiTop

, order no. 209 012). Ensure the batteries are inserted the right

way round!

- Insert the upper case with the lug in the locking pin (lower case).

Caution! Do not crush the cable connection!

- Close the upper case (let the hook snap into place).

Check if the battery change was successful:
“Batteries: OK“ must appear in “GLP/Tools - Check -

info”.


The batteries have no effect on the data integrity of the OxiTop

®

-C

Disposing of the batteries

Properly dispose of used batteries at a battery collection point.