GLP / Tools

GLP / Tools

70

BA31114e5 07/2006

BA31114e5 07/2006

71

The controller sets a new test
date in the measuring head. The
controller calculates the new test
date from the current date + the
check interval set up (see the
chapter “GLP/Tools - Settings -
GLP - Check interval”).

The controller displays the setting
of the calibration date.





If the date was not set successfully (e.g. because the controller was
not held to - or not held close enough to - the measuring head):

Repeat the procedure using

,

then continue as described
above.



Pneumatic test

The pneumatic test tests the measurement precision of the measuring head. It says
nothing about the long-term impermeability of the system.

To perform the pneumatic test, you require the testing agent, OxiTop

®

PT (WTW

order number 209 334).

The user interface of the controller guides you through the test:


“Check“ menu.

Use

to select the

menu item “Pneumatic test

“.

Set the plunger of the siringe on
the OxiTop

®

PT testing agent to

5 scale parts.
Tightly screw the measuring
head to be tested onto the
OxiTop PT tesing agent.


(Exceeding the time causes a
return to the “Check“ menu“).