Xylem OxiTopIS OC100 User Manual

Page 59

Advertising
background image

GLP / Tools

GLP / Tools

58

BA31114e5 07/2006

BA31114e5 07/2006

59


Set each of the number blocks
underlaid in black using

.

Confirm and continue with

.

Set the day, month, year, hours,
minutes, seconds consecutively.

The measuring time of samples already started is not affected by a change of date
and time.

GLP

The GLP (Good Laboratory Practice) set of rules presents particular requirements
concerning test resource monitoring for the measuring systems implemented
(quality assurance).
The following points must be fulfilled to comply with the requirements of GLP,:

1. Documentation of the calibration intervals.


2. Reporting of the calibrations:

Identification no. of the measuring head/controller

Date and signature of the operator.


3. Determination of the calibration frequency and instructions for the
performance
:

The frequency depends on the operating conditions and the data of the test plan.


The OxiTop

®

Control system supports these standards using an add-on GLP mode:

This mode inhibits measuring if the calibrating interval is exceeded.
As a result of these restrictions, the GLP mode effects a standardization of the
calibrating and measuring conditions that are indispensable for working from the
aspect of quality assurance.

Switch the GLP function on or off (only in the standard BOD operating mode):


“Settings“ submenu:

Use

to select the “GLP“ submenu.


“GLP“ submenu.
“GLP“ is preselected.

Press

,

then use

to select

either GLP “On“ or “Off“

Confirm using

.




GLP - calibrating interval (“Calinterval”)

Set the time period (1 to 36 months) here. When it expires, the instrument registers
that the next test resource monitoring of the measuring heads is due. After the
calibrating interval expires, measuring is blocked until the calibration is performed or
the GLP setting is set to “Off“.

If the end of the calibrating interval set up lies within the measuring time of a
measurement to be started, this measurement cannot be started.

“GLP“ submenu.

Use

to select the

subitem, “Calinterval“.