Operating Mode: Standard BOD
Operating Mode: Standard BOD
46
BA31114e5 07/2006
BA31114e5 07/2006
47
Excluding a curve
This function is used to temporarily exclude a single curve (e.g. an outlier) from the
evaluation and averaging of a whole sample.
The curve is only excluded temporarily! The data stock of the sample
management does not change. The excluded curve is present again when the call is
repeated.
The function “List of measuring heads“ (see the chapter “Sample management”) is
used to find leaky or defective single samples.
Display of the single curves:
Select curve.
:
Change to
“Exclude curve“.
“Exclude curve“
is preselected.
“Return“: Returns to the previous
menu.
Message
“Curve excluded“.
Updated display
(curve excluded, mean value
newly calculated).
Printout of the updated results
with curves
(without the excluded curve).