What to do if...?

What to do if...?

90

BA31114e5 07/2006

BA31114e5 07/2006

91

Requirements /
Problems

Procedure / Problem solutions

Measuring head
unintentionally started
for measurement

Measuring head
started with incorrect
settings

Measuring head is
required for another
sample

Problem solution
The measuring head can be released again through the
command Reset/release:
Chapter “GLP/Tools

-

Maintenance - Reset/release

.

The controller guides you further through the operation.

If only one measuring head is stored under the
corresponding sample number (in the routine BOD
operating mode, this is always the case), the controller
automatically erases the sample in the sample
management.

Search for the
defective measuring
head

Procedure:
• Perform a measuring head reaction test (see above).

A measuring head that repetitively shows no reaction to
the test is defective.

• If the defective measuring head cannot be established

in this way, perform the measuring head check (see
“Single check”). In doing this, each individual measuring
head must be checked until the defective measuring
head has been found.

A measuring head is
missing or is defective.

Requirement:
To determine the
corresponding sample

Procedure:
Call up the function “Call up data” from the sample
management for each individual unfinished sample. The
corresponding sample can be determined by this.
For subsequent error handling, see the chapter “Sample
management - Call up data”.

Requirements /
Problems

Procedure / Problem solutions

The precision and
sealing of the system
sample bottles plus
measuring head
should be checked

Procedure:

See the chapter “GLP/Tools

-

Check - Cal test “

Measuring head check
(single check)

(To which running
sample does the
selected measuring
head belong?)

Procedure:
Single check method:

menu “GLP/Tools

-

Check -

info”

• The controller guides you further through the operation.
• All single information on the state of the measuring

heads is listed, i.e. the measuring head responds.

• The check can be undertaken at any time without

affecting the measurement that is running.

• If no reaction at all can be determined, the batteries

should be changed and the measuring head check
repeated.

• The measuring precision cannot be checked with this!

The measuring
precision of the
measuring head
should be checked.

Procedure:
Pneumatic test (PT) of the measuring head:

see the chapter “GLP/Tools

-

Check - Pneumatic test”

Perform measuring
head reaction test

Procedure:
• Menu “GLP/Tools

-

Check - Show all

:

• All optically attainable measuring heads must flash for

approx. 5 seconds independent of their status.

• The check can be undertaken at any time without

affecting the measurement that is running.

• This test only checks the reaction to commands.
• If no measuring head reacts, the controller should be

checked.