Technical Data
Technical Data
96
BA31114e5 07/2006
BA31114e5 07/2006
97
OxiTop
®
-C measuring head
Measuring principle
manometric by means of a pressure measuring head
Measured of
BOD
n
(oxygen drift according to DIN 38409 T 52)
Pressure range
500 ... 1350 hPa (mbar).
Accuracy
± 1% of measured value ± 0.7 % of the measuring
range
Resolution
1 hPa
Display
LED pilot lamps
Power supply
2 x lithium batteries, CR2430, e.g. WTW type
Batt/OxiTop
, order no. 209 012)
Battery run time
Typically 2 years, in case of normal use
Safety class
3, equipment safety: IEC 1010, EN 61010 part 1
Protection type
IP 54 DIN 40050
EMC
-Interference emission
-interference immunity
EN 50081-1, FCC Class A
EN 50082-1
Climate class
2, VDI/VDE 3540
Ambient temperature
Storage: -25°C ... +65°C
Operation: +5°C ... +50°C
Dimensions
H: 70 mm,
∅
∅
∅
∅
70 mm
Weight
95 g
Test certificates
CE, TÜV, UL, CUL(UL/CSA)
Warranty