Operating Mode: Standard BOD

Operating Mode: Standard BOD

50

BA31114e5 07/2006

BA31114e5 07/2006

51

Measured values outside the measuring range


In the following cases, “undef.“ (undefined) is displayed instead of the measured
value or mean value during evaluation:
- A measured curve exceeding the measuring range at any point of its path.
- A measured curve undercutting the measuring range at its end point.

After the exclusion of the defective curve (see chapter “Exclude curve“), the controller
displays the mean value again.

Sample displays:


A measured value exceeding the
measuring range (Overflow).


A measured value undercuts the
measuring range (Underflow).


A measured value temporarily
leaves the measuring range.

Print

Switch on the TD100 IR printer.

Point the controller at the printer:

10 cm to 100 cm

The controller outputs the report data of the selected sample to
the IR interface.
Printout of
- sample statistics (mean value, SD, n)
- sample information (type, running time, measuring range, start
date and start time, end date, identification no.)
- curve + BOD final value
- with GLP On: additional information on the system status
and list of the series nos. of the measuring heads allocated to
the sample
- If AutoTemp is switched on: note “AutoTemp on“
- Report printouts:

- current settings

- start information

-

info

- controller info

From the sample management:

From the evaluation (display of
all the curves):