Technical Data

Technical Data

96

BA31114e5 07/2006

BA31114e5 07/2006

97

OxiTop

®

-C measuring head


Measuring principle

manometric by means of a pressure measuring head

Measured of

BOD

n

(oxygen drift according to DIN 38409 T 52)

Pressure range

500 ... 1350 hPa (mbar).

Accuracy

± 1% of measured value ± 0.7 % of the measuring
range

Resolution

1 hPa

Display

LED pilot lamps

Power supply

2 x lithium batteries, CR2430, e.g. WTW type
Batt/OxiTop

, order no. 209 012)

Battery run time

Typically 2 years, in case of normal use

Safety class

3, equipment safety: IEC 1010, EN 61010 part 1

Protection type

IP 54 DIN 40050

EMC
-Interference emission
-interference immunity


EN 50081-1, FCC Class A
EN 50082-1

Climate class

2, VDI/VDE 3540

Ambient temperature

Storage: -25°C ... +65°C
Operation: +5°C ... +50°C

Dimensions

H: 70 mm,

70 mm

Weight

95 g

Test certificates

CE, TÜV, UL, CUL(UL/CSA)

Warranty

1 year