Operating Mode: Standard BOD

28

BA31114e5 07/2006

BA31114e5 07/2006

29

Operating mode: Standard BOD

Sample preparation

See WTW application reports
(contained within the scope of delivery of the accessories supplied).


Screw the OxiTop

-C measuring heads onto the sample bottles

and close them tightly.

Important note:
Never use joint grease or other lubricant for your OxiTop

-C measuring

heads. Some of these products contain solvents that can cause severe damage
to the plastic housing.
The sealing of the BOD bottles is also perfectly adequate without grease.
However, you should always wipe off heavy contamination and particles on the
sealing surfaces of the rubber sleeves and OxiTop

-C. WTW accepts no

liability for damage due to the use of joint grease.

Start the measurement


To change to the operating mode BOD Standard refer to chapter GLP/Tools -
Settings - Operating mode.

“Communication with the
measuring heads” mode.

Preselected: “Start sample”.

Use

to select the measurement range.
The filling volume required is
given in the right hand column.


The controller stores the setting
(memory function: the last
selected measurement range is
set).