Xylem OxiTopIS OC100 User Manual

Page 40

Advertising
background image

Operating Mode: Standard BOD

Operating Mode: Standard BOD

40

BA31114e5 07/2006

BA31114e5 07/2006

41


Jumps to submenu of the
selected sample.
Preselected: “Call up data“

Point the controller at the measuring
heads:

40 cm to 100 cm

Sequential call up of the
measuring heads allocated to the
sample (in this example: 7) in the
scanner mode:
The controller
- stores the data obtained
- updates the list of the
sample management

Duration of a pass:
approx. 3 seconds for 12
measuring heads in a stirring
system.

The controller confirms the call
up performed.

The controller then returns
automatically to the submenu.

If the measuring heads do not all respond, the following display appears:

Renewed start of the data call up.

The controller continues to search for missing measuring heads in the scanner mode.
Without pressing any further keys, the data call up can be continued at other
locations (e.g. other stirring platforms or other thermal cabinets). Approximately 7
seconds after the last request was performed, the scanner mode automatically breaks
off and the following display appears:

Use

to restart the

the data transfer (see above).

Return to the main menu,
“Sample management“.
Transferred and completed
samples appear with a tick next
to them. (The complete
measurement data records of all
finished measuring heads of the
sample have been procured.)

Immediately after the successful data transfer of a sample, the allocated measuring
heads are given the “free” status. The measuring heads can be used for a new
measurement.

Advertising