Xylem OxiTopIS OC100 User Manual

Page 66

Advertising
background image

GLP / Tools

GLP / Tools

66

BA31114e5 07/2006

BA31114e5 07/2006

67

info

“Check“ menu.

Use

to select “

info

.

The measuring head flashes and
a display appears on the
controller giving the following
information:

- serial number of the measuring
head,
- battery status
(OK/LOBAT/EMPTY!!),
- the next calibration date
(only in the measuring mode,
Standard BOD and GLP “On“
- status of the measuring head
(free/used/defective).

If the measuring head is “used“,

use

to scroll through the

display of further information:
- start date
- final date
- sample number
- type and
- range for the measurement


Return to the “Check“ menu

using

.

Repeat the procedure for each
measuring head.

If the measuring head does not respond, the following display
appears after approximately 7 seconds:

The query was stopped.
You can select between
- continue query (see above) and
- Stop (return to the
“Check“ menu

Controller info

“Check“ menu.

Use

to select

“Controller info“.

The display shows the following
information:
- number of measuring heads for
which there is still memory
capacity
- status of the supply batteries
- status of the data backup
batteries
(OK/LOBAT).

Use

to scroll to the

specification of the
- software version
- Ser.no.
(can differ from the example
shown here)


Return to the “Check“ menu

using

.