DESCRIPTION
The TIME_DIV command modifies the
timebase setting. The new timebase setting may
be specified with suffixes: NS for nanoseconds,
US for microseconds, MS for milliseconds, S
for seconds, or KS for kiloseconds. An out-of-
range value causes the VAB bit (bit 2) in the
STB register to be set.
The TIME_DIV? query returns the current
timebase setting.
COMMAND SYNTAX
Time_DIV <value>
<value>:={1NS(not every type has this
value),2.5NS,5NS,10NS,25NS,50NS,100NS,25
0NS,500NS,1US,2.5US,5US,10US,25US,50US,
100US,250US,500US,1MS,2.5MS,5MS,10MS,
25MS,50MS,100MS,250MS,500MS,1S,2.5S,5S
,10S,25S,50S}
QUERY SYNTAX
Time_DIV?
RESPONSE FORMAT
Time_DIV <value>
EXAMPLE
The following sets the time base to 500 μs /div:
Command message:
TDIV 500US
RELATED COMMANDS
TRIG_DELAY, TRIG_MODE
ACQUISITION
TIME_DIV, TDIV
Command /Query