DESCRIPTION
The OFFSET command allows adjustment of
the vertical offset of the specified input
channel. The maximum ranges depend on the
fixed sensitivity setting.
If an out-of-range value is entered, the
oscilloscope is set to the closest possible
value and the VAB bit (bit 2) in the STB
register is set.
The OFFSET? query returns the offset value
of the specified channel.
COMMAND SYNTAX
<channel>: OFfSeT <offset>
<channel> : = {C1, C2, C3,C4}
<offset> : = See specifications.
QUERY SYNTAX
<channel>: OFfSeT?
RESPONSE FORMAT
<channel>: OFfSeT <offset>
EXAMPLE
The following command sets the offset of
Channel 2 to -3 V:
Command message:
C2: OFST -3V
ACQUISITION
OFFSET, OFST
Command /Query