B&K Precision 2552-2559 - Programming Manual User Manual
Page 119

119
SR,< source > :is used to set the trigger’s
channel.If you want to set the other option.
You must set it.
<source>: = {C1, C2, C3,C4,EX, EX5}
CHAR, <characteristicse>:is used to set the
standard .if you want to set it, the <trig_type>
must be set to TV.
<characteristicse>:={NTSC, PALSEC}
SYNC,<sync_type>:is used to set sync. If you
Want to set it. You must set <trig_type> to TV
<sync_type> : = {AL,LN,OF,EF}
AL means all lines; LN means line num; OF
means odd field; EF means even field
LINE,<line>:is used to set the line num. if you
want to set it. The SYNC must be set to
LINENUM
POL,<polarity>: is used to set polarity. If you
want to set it. You must set <trig_type> to TV
<polarity>: = {PO,NE}
PO means positive. NE means negative.
VERT,<vertical>: is used to set vertical. If you
Want to set it. You must set <trig_type> INTV
<vertical>: = {UP,DOWN,BOTH}
QUERY SYNTAX
TRig_SelEct?
RESPONSE FORMAT
TRig_SelEct <mode>,the other options
EXAMPLE
The following sets the trigger type to video, the
trigger source to C1,the standard to NTSC, the
polarity to positive, the sync to line num and the
line num to 5:
TRSE TV,SR,C1,CHAR,NTSC,POL,
PO,SYNC,LN,LINE,5
RELATED COMMANDS
TRSL VTCL