B&K Precision 2552-2559 - Programming Manual User Manual
Page 116

116
DESCRIPTION
The TRIG_LEVEL command adjusts the trigger
level of the specified trigger source. An out-of-
range value will be adjusted to the closest legal
value and will cause the VAB bit (bit 2) in the
STB register to be set.
The TRIG_LEVEL? query returns the current
trigger level.
COMMAND SYNTAX
<trig_source>: TRig_LeVel <trig_level>
<trig_source>: = {C1, C2, C3, C4, EX, EX5}
<trig_level>: = -6DIV* volt/div to 6DIV *
volt/div
Note: The suffix V is optional and assumed.
QUERY SYNTAX
<trig_source>: TRig_LeVel?
RESPONSE FORMAT
<trig_source>: TRig_LeVel <trig_level>
EXAMPLE
The following code adjusts the trigger level of
Channel 3 to 52.00mv:
Command message:
C3:TRig_LeVel 52.00mv
RELATED COMMANDS
TRIG_COUPLING, TRIG_DELAY,
TRIG_MODE, TRIG_SELECT, TRIG_SLOPE
ACQUISITION
TRIG_LEVEL, TRLV
Command /Query