DESCRIPTION
The *ESE command sets the Standard Event
Status Enable register (ESE). This command
allows one or more events in the ESR register
to be reflected in the ESB summary message
bit (bit 5) of the STB register.
COMMAND SYNTAX
*ESE <value>
<value> : = 0 to 255
QUERY SYNTAX
*ESE?
RESPONSE FORMAT
*ESE <value>
EXAMPLE
The following instruction allows the ESB bit to
be set if a user request (URQ bit 6, i.e.
decimal 64) and/or a device dependent error
(DDE bit 3, i.e. decimal 8) occurs. Summing
these values yields the ESE register mask
64+8=72.
Command message:
*ESE 72