Section 5

Operation

The AMTAX sc can only be operated with an sc1000 controller. For
further information, see the sc1000 User Manual.

An LED on the door indicates the current operating state. Refer to
the sc1000 User Manual and

section 7.2.1, page 57

.

Instrument, chemicals and electrode are temperature-sensitive. To
prevent incorrect measurements, only operate the instrument with
the door closed.

5.1 Sensor diagnostics menu

SELECT AMTAX sc (if more than one sensor or analyzer is attached)

AMTAX sc

ERROR LIST

Displays all errors currently present in the sensor

WARNING LIST

Displays all warnings currently present in the sensor

5.2 Sensor setup menu

SELECT AMTAX sc (if more than one sensor or analyzer is attached)
CALIBRATION (see

5.3, page 42

)

CORR. FACTOR

Displays the correction factor

LOCATION 1

Displays location1 which is adjusted in the CONFIGURE menu

GAIN CORR

Adjusts the correction factor for channel 1

LOCATION2

Available with the 2-channel version.

GAIN CORR

Available with the 2-channel version.

CALIBRATE

Triggers an automatic calibration, then measuring mode

CAL.-CLEAN

Triggers an automatic calibration with subsequent automatic cleaning, then measuring mode.

SETTINGS

SET OUTMODE

Value that is output during a calibration and the following discarded values.
HOLD = last measured value, SET TRANSFER = value to be entered.

AUTOCAL

SET INTERVAL

Time interval between two calibrations

START

Start time for the calibration (In case of more than one calibration per day: Enter start time for
the first calibration.)

DISCHARGE

Number of measured values that are discarded after a calibration.

SEL ADJ METHOD

Select the calibration method

ADVANCED

Default calibration method for software version

≥ 1.60 (For best accuracy with lowest

measuring values.)

CONVENTIONAL

Calibration method for software version

< 1.60 is available for compatiblility reasons (Not

selectable in the measuring range 0.02–5 mg/L.).

DEFAULT SETUP

Resets to factory default settings.

CONFIGURE

LOCATION 1

Settings for location 1

EDIT NAME

Enter the name for the measuring location as required.

SET PARAMETER

Select output: ammonium or ammonium nitrogen

SELECT UNITS

Select output: mg/L or ppm

QUANTITY CH 1

Number of measurements in succession ( = measurements on channel 1 +
DISCHARGE VAL 1 channel 1). Available with the 2-channel version.