Hach-Lange SI794 D User Manual User Manual

Page 41

Advertising
background image

°C

(°F)

Code Display

Action

Choices

o2.

Specify temperature unit

Select using key key.

Press enter to proceed.

Select temperature probe

Select using key key.

Press enter to proceed.

22NTC

(30NTC)

4 - 20 mA

0 - 20 mA

Set output current range

Select using key key.

Press enter to proceed.

Current start: Enter lower end of

scale.

Select position using , edit number

using , press enter to proceed.

000.0 °C

Current end: Enter upper end of

scale.

Select position using , edit number

using , press enter to proceed.

100.0 °C

Output current

Process

temperature

20

4

100

0

Process temperature: Current start and current end

Process

temperature

20

4

70

50

[mA]

Example 1: Range 0 ... 100 °C

Example 2: Range 50 ... 70 °C.

Advantage: Higher resolution in

range of interest

[°C]

[°C]

[mA]

Output current

41