Hach-Lange SI794 D User Manual User Manual

Page 69

Advertising
background image

Display

Action

Remark

Enter default for concentration
(Select position using

,

edit number using

,

press enter to confirm.)

Display of new slope and

zero point (related to 25 °C at

1013 mbars).

Press enter to end

concentration calibration.

The default value

is calculated from

rel. humidity,

cal pressure, and

cal temperature.

(the unit of

measurement,

ppm or mg/L, ... ,

is preset during

configuration.)

Place sensor in process

The new value is shown in the

main display alternately with

“Hold”; “enter” blinks.

Press enter to exit calibration.

After end of

calibration, the out-

puts remain in Hold

mode for approx.

20 sec.

New calibration:

Press cal key.

Information on concentration calibration (Conc)

Calibration in air. This calibration method is recommended when the sensor can be

removed for calibration. Air has a stable oxygen content. Therefore the adjustment

processes during calibration run more quickly.
For 2-point calibration, perform zero calibration first.

69