Display

Action

Remark

Enter setpoint value for

saturation

Select position using ,

edit number using ,

press enter to proceed.

Display of new slope and

zero point (related to 25 °C at

1013 mbars)

Press enter to end calibration.

Default: last value

entered

Place sensor in process.

The percent saturation is

shown in the main display

alternately with “Hold”;

“enter“ blinks.

Press enter to exit calibration.

After end of

calibration, the out-

puts remain in Hold

mode for approx.

20 sec.

New calibration:

Press cal key.

Information on saturation calibration (SAT)

The calibration medium must be in equilibrium with air (percent saturation for

water is 100 %). Oxygen exchange between water and air is very slow. To speed

up the adjustment processes, make sure that there is a steady medium flow

during calibration.

If the percent saturation is known from a simultaneous measurement, it can be

entered manually.

For 2-point calibration, perform zero calibration first.