Product calibration

Calibration by sampling

Display

Action

Remark

Product calibration step 1:

Select calibration

(Press cal key).

Enter passcode: 1105
(Select position using

,

edit number using ,

press enter to confirm.)

Save value.

Press enter to proceed.

The type of product

calibration (SAT or

Conc) is selected during

configuration (mea-

surement procedure).

If an invalid code is

entered, the transmitter

returns to measuring

mode.

Now the reference

value must be

determined.

The transmitter

goes to measuring

mode.

During product calibration the sensor remains in the process.

The measurement process is only interrupted briefly.

Procedure: During sampling the currently measured value is saved in the

transmitter. The transmitter immediately returns to measuring mode.

The calibration mode indicator blinks and reminds you that calibration

has not been terminated. The comparison value is measured on the site,

e.g. using a portable DO meter in a bypass. This value is then entered in

the transmitter. The new value for slope or zero is calculated from the

stored value and the comparison value. From the measured value, the

transmitter automatically recognizes whether a new slope or zero must

be calculated (above approx. 5 % saturation: slope, below: zero).

If the sample is invalid, you can take over the measured value saved

during sampling instead of the comparison value. In that case the old

calibration values remain stored. Afterwards, you can start a new product

calibration.

The following describes a product calibration with slope correction – a

product calibration with zero correction is performed correspondingly.