Error messages, Error codes), Error display problem possible causes – Hach-Lange SI794 D User Manual User Manual

Page 82

Advertising
background image

Alarm contact

Red LED

Out 1 (22 mA)

Out 2 (22 mA)

Error messages

(error codes)

SAT range

Sensor defective

Wrong sensor connected

Measuring range exceeded

Error

Display Problem

Possible causes

ERR 01

Measured

value

blinks

x

x

x

Conc range

Sensor defective

Wrong sensor connected

Measuring range exceeded

ERR 02

Measured

value

blinks

x

x

x

System error

Configuration or calibration data

defective; completely reconfigure

and recalibrate the transmitter.

Memory error in device program

(PROM defective)

ERR 98

“ConF”

blinks

x

x

x

x

Factory settings

EEPROM or RAM defective

This error message only occurs

in the case of a total defect. The

transmitter must be repaired and

recalibrated at the factory.

ERR 99

“FAIL”

blinks

x

x

x

x

82