Diagnostics functions – Hach-Lange SI794 D User Manual User Manual
Page 75

Display of output currents
Press enter while in measuring mode.
The current at output 1 is shown in the main display,
the current at output 2 in the secondary display.
After 5 sec the transmitter returns to measuring
mode.
Diagnostics functions
Display
Remark
Display of calibration data (Cal Info)
Press cal while in measuring mode and enter code:
0000. The slope is shown in the main display, the zero
current in the secondary display.
After 20 sec the transmitter returns to measuring
mode (immediate return at pressing enter).
Display of sensor current (sensor monitor)
Press conf while in measuring mode and enter code:
2222. The (uncompensated) sensor current is shown
in the main display, the measuring temperature in
the secondary display.
Press enter to return to measurement.
Display of last error message (Error Info)
Press conf while in measuring mode and enter code:
0000. The last error message is displayed for approx.
20 sec.
After that the message will be deleted (immediate
return to measurement at pressing enter).
75