Hach-Lange SI794 D User Manual User Manual

Page 51

Advertising
background image

LiMIT

(CtROL)

Code Display

Action

Choices

rL.

Use of relays:

• Limit function (LiMIT)

• Controller (CtROL)

Select using ,

press enter to proceed.

Note: Selecting CtROL leads to

Controller menu group Ct.

Lo

(Hi)

For limit 1 function principle,

see Fig. on Pg. 53. Select using ,

press enter to proceed.

Limit 1 contact response

N/C: normally closed contact

N/O: normally open contact
Select using .

Press enter to proceed

N/C

(N/O)

Limit 1 setpoint

Select using , edit number

using , press enter to proceed.

0000 %

(

xxxx %)

Limit 1 hysteresis

Select using , edit number

using , press enter to proceed.

0001 %

(

xxxx %)

Limit 1 delay

The contact is activated with

delay (deactivated without delay)

Select position using ,

edit number using ,

press enter to proceed.

0010 sec

(

0 ... 600 sec)

L1.

51