Alarm contact

Relay contact, floating

Contact rating

AC < 250 V / < 3 A / < 750 VA

DC < 30 V / < 3 A / < 90 W

Contact response

N/C (fail-safe type)

Response delay *

0000 ... 0600 sec

Limit values

Output via relay contacts R1, R2

Contacts R1, R2, floating but inter-connected

Contact rating

AC < 250 V / < 3 A / < 750 VA

DC < 30 V / < 3 A / < 90 W

Contact response *

N/C or N/O

Response delay *

0000 ... 9999 s

Setpoints *

Within selected range

Hysteresis *

000.0 ... 050.0 % / 00.00 ... 05.00 mg/L (ppm)

PID process controller Output via relay contacts R1, R2

(Relay R1: below setpoint,

Relay R2: above setpoint)

Setpoint specification * 0 ... 500 % / 0.00 ... 50.00 mg/L (ppm)

Neutral zone *

000.0 ... 050.0 % / 00.00 ... 05.00 mg/L (ppm)

P-action component * Controller gain Kp:

0010 ... 9999 %

I-action component * Reset time Tr:

0000 ... 9999 s

(0000 s = no integral action)

D-action component * Rate time Td:

0000 ... 9999 s

(0000 s = no derivative action)

Controller type *

Pulse length controller or pulse frequency controller

Pulse period *

0001 ... 0600 s, min. ON time 0.5 s

(pulse length controller)

Max. pulse frequency * 0001 ... 0180 min

-1

(pulse frequency controller)