Hach-Lange SI794 D User Manual User Manual

Page 87

Advertising
background image

Type B sensor (Hamilton Oxygold)

Thermometer and Sensoface:

Temperature out of concentration or saturation range

Slope

Zero point

Response

time

Cal timer

Range

200 ... 550 nA

-2 ... +2 nA

max. 720 sec

> 250 ... < 500 nA

> - 0.5 ... < 0.5 nA

< 300 s

< 80 %

225 ... 250 nA

or

500 ... 525 nA

-1.0 ... -0.5 nA

or

+0.5 ... +1.0 nA

300 ... 600 s

80 ... ≤ 100 %

< 225 nA or

> 525 nA

< -1.0 nA or

> +1.0 nA

> 600 sec

Timer expired

Sensocheck

Continuously monitors the sensor and leads for short circuits or

open circuits. Critical values make the Sensoface “sad” and the

corresponding icon blinks:

The Sensocheck message is also output as error message Err 33.

The alarm contact is active, the red LED is lit, output current 1 is

set to 22 mA (when configured correspondingly). Sensocheck

can be switched off during configuration (then Sensoface is also

disabled).

Exception: After a calibration a smiley is always displayed for

confirmation.

87