Configuration

Calibration mode

Alarm settings

Calibration mode

Cal timer interval

Select Sensocheck

Delay

LED in HOLD mode

Code Display

Menu group

Select menu item

Calibration mode

CA.

Alarm settings

AL.

Exit:

Press conf, then enter.

15

16

Alarm

Alarm contact

The alarm contact is closed during normal operation (N/C).

It opens in the case of alarm or power outage.

As a result, a failure message is provided even in the case

of line breakage (fail-safe behavior). For contact ratings,

see Specifications.

Error messages can also be signaled by a 22 mA

output current (see Pg 39,

The operating behavior of the alarm contact is shown

on Pg 84.

Exit:

Press conf, then enter.

The alarm delay acts on the LED, the 22 mA signal, and the

alarm contact.

45, 83).