Hach-Lange SI794 D User Manual User Manual

Page 71

Advertising
background image

Display

Action

Remark

Main display:

Zero current

Press enter to save the value

or correct it using the arrow

keys and then save by

pressing enter.

Secondary display:

Sensor current measured

Display of slope

Display of new zero current.

Press enter to end calibration,

place sensor in the process

New calibration:

Press cal key.

Select calibration

(Press cal key).

Enter passcode: 1001
Select position using ,

edit number using ,

press enter to proceed.

Zero

Place sensor in oxygen-free

medium

The oxygen value is shown in

the main display alternately

with “Hold”, “enter” blinks.

Press enter to stop Hold.

After end of

calibration, the out-

puts remain in Hold

mode for approx.

20 sec.

The transmitter is in

the Hold mode.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.