Calibration to percent saturation (sat) – Hach-Lange SI794 D User Manual User Manual

Page 66

Advertising
background image

Display

Action

Remark

Enter relative humidity

Select position using ,

edit number using ,

press enter to proceed.

Enter calibration pressure

Select position using , edit

number using , press enter

to proceed.

Default for calibra-

tion pressure is the

process pressure

configured

Calibration to percent saturation (SAT)

Select calibration

(Press cal.)

Enter passcode: 1100

Select position using ,

edit number using ,

press enter to proceed.

Place sensor in calibration

medium

Start with enter

Automatic drift check

Display of sensor current (re-

lated to 25 °C and 1013 mbars)

and measuring temperature.

The drift check might take

some time.

Drift check can

be stopped after

> 10 sec by press-

ing cal (accuracy

reduced).

SAT or Conc calibration

is selected during

configuration.

If an invalid code is en-

tered, the transmitter

returns to measuring

mode.

Default for relative

humidity in aqueous

media:

rH = 100 %

(in air approx. 50 %)

The transmitter is in

Hold mode

66