Calibration to concentration (conc) – Hach-Lange SI794 D User Manual User Manual

Page 68

Advertising
background image

Display

Action

Remark

Enter relative humidity
(Select position using

,

edit number using

,

press enter to confirm.)

Enter calibration pressure
(Select position using

,

edit number using

,

press enter to confirm.)

Default for calibra-

tion pressure is

normal pressure

1.013 bars.

Calibration to concentration (Conc)

Select calibration

(Press cal.)

Enter passcode: 1100
(Select position using

,

edit number using

,

press enter to confirm.)

Place sensor in air

Start with enter

Automatic drift check

Display of input current (re-

lated to 25 °C and 1013 mbars)

and measuring temperature.

The drift check might take

some time.

Drift check can be

stopped after

> 10 sec by press-

ing cal (accuracy

reduced).

SAT or Conc calibration

is selected during

configuration.

If an invalid code is

entered, the transmitter

returns to measuring

mode.

Default for relative

humidity in air:

rH = 50 %

The transmitter

goes to HOLD

mode.